SMART: A Swing-Assisted Multiplexed Analyzer for Point-of-Care Respiratory Tract Infection Testing

Li Zhang, Xu Wang, Dongchen Liu, Yu Wu, Li Feng, Chunyan Han, Jiajia Liu, Ying Lu, Dmitriy V. Sotnikov, Youchun Xu, Jing Cheng

Journal:Biosensors-Basel

IF:5.4

DOI:10.3390/bios13020228

PMID:36831994

Published:2023-02-04

research field:

Abstract

Respiratory tract infections such as the ongoing coronavirus disease 2019 (COVID-19) has seriously threatened public health in the last decades. The experience of fighting against the epidemic highlights the importance of user-friendly and accessible point-of-care systems for nucleic acid (NA) detection. To realize low-cost and multiplexed point-of-care NA detection, a swing-assisted multiplexed analyzer for point-of-care respiratory tract infection testing (SMART) was proposed to detect multiple respiratory tract pathogens using visible loop-mediated isothermal amplification. By performing hand-swing movements to generate acceleration force to distribute samples into reaction chambers, the design of the SMART system was greatly simplified. By using different format of chips and integrating into a suitcase, this system can be applied to on-site multitarget and multi-sample testing. Three targets including the N and Orf genes of SARS-CoV-2 and the internal control were simultaneously analyzed (limit of detection: 2000 copies/mL for raw sample; 200 copies/mL for extracted sample). Twenty-three clinical samples with eight types of respiratory bacteria and twelve COVID-19 clinical samples were successfully detected. These results indicate that the SMART system has the potential to be further developed as a versatile tool in the diagnosis of respiratory tract infection.Keywords:LAMP;point-of-care;SARS-CoV-2;microfluidics;multiplexed detection

本文使用的Yeasen产品

购物车
客服
转染试用